EUROPE


Standard Range

i-series

Standard Type i220/SA1AM

Universally applicable with a sales success for over 15 years.
The i-series is a standard range and easier to maintain than customer tailored products.

Data Sheet

Data Sheet

Data Sheet DOWN LOAD

Features

Features

■ Test range:

・Maximum acceleration: 1250 m/s2

・Maximum velocity: 3.5 m/s

・Maximum displacement: 51 mmp-p

・Maximum loading mass: 200 kg

■ Patented upper (armature) support system PS Guide

Parallel Slope Guide is standard

■ All models can be directly coupled to a climatic chamber.

Table Insert Pattern (Unit:mm)

Table Insert Pattern(Unit:mm)

Table Insert Pattern Table Insert Pattern

Specifications

Ecology

  System Model i210/SA1M i220/SA1M i220/EM1AM
  Data Sheet
  Frequency Range (Hz) 0-4000 0-3300 0-3300
System
Specifications
Rated
Force
Sine (kN) 3 8 8
Random (kN rms)*1 3 8 8
Shock (kN) 9 16 16
High Velocity Shock*5(kN) - - 10
Maximum
Acc.
Sine (m/s2) 1000 1250 1250
Random (m/s2 rms) 700 875 875
Shock (m/s2 peak) 2000 2000 2000
High Velocity Shock*5(m/s2 peak) - - 1562
Maximum
Vel.
Sine (m/s) 2.2 2.2 2.2
Shock (m/s peak) 2.2 2.2 2.2
High Velocity Shock*5(m/s peak) - - 3.5
Maximum
Dsp.
Sine (mmp-p) 30 51 51
High Velocity Shock*5(mmp-p) - - 51
Maximum Travel (mmp-p) 40 60 60
Maximum Load (kg) 120 200 200
Power Requirements (kVA)*2 6.8 16.4 16.4
Breaker Capacity (A)*3 15 30 30
Vibration
Generator
Model i210 i220 i220
Armature Mass (kg) 3 6.4 6.4
Armature Diameter (φmm) 128 190 190
Allowable Eccentric Moment (N・m) 160 294 294
Dimensions (mm) W×H×D 868 × 700 × 458 1020 × 903 × 550 1020 × 903 × 550
Shaker Body Diameter 458 550 550
Mass (kg) 350 900 900
Power
Amplifier
Model SA1M-i10 SA1AM-i20 EM1AM-i20
Maximum Output (kVA) 5 10 10
Dimensions (mm) W×H×D 580 × 1950 × 850 580 × 1950 × 850 580 × 1950 × 850
Mass (kg) 240 280 280
Controller Vibration Controller See Vibration Controller K2
Cooling Cooling Method Air cooling
Blower Dimensions (mm) W×H×D*4 600 × 1905 × 557 808 × 2085 × 733 808 × 2085 × 733
Mass (kg) 45 85 85
Wattage (kw) 0.4 1.5 1.5
Duct Hose Diameter (φ) 125 125 125
  • *1 Random force ratings are specified in accordance with ISO5344 conditions. Please contact IMV or your local distributor with specific test requirements.
  • *2 Power supply: 3-phase 380/400/415/440 V, 50/60 Hz. A transformer is required for other supply voltages.
  • *3 Breaker capacity for 400 V
  • *4 Spec described above is in case of 50Hz. Dimension is changed in case of 60Hz.
  • *5 In case of high velocity option
  • * The specification shows the maximum system performance. For long-duration tests, de-rating by up to 70 % must be applied. Continuous use at maximum levels may cause failure. Please contact IMV if you use more than 70 %.
  • * In the case of Random vibration test, please set the test definition of the peak value of acceleration waveform to be operated less than the maximum acceleration of Shock.
  • * Frequency range values vary according to sensor and vibration controller.
  • * Armature mass and acceleration may change when chamber is comibined.